Fujifilm Medical Systems only duodenoscope manufacture to have its revised reprocessing instructions validated by the FDA
Accreditation Insider, January 19, 2016
Want to receive articles like this one in your inbox? Subscribe to Accreditation Insider!
In October, the Food and Drug Administration (FDA) ordered the top three manufacturers of duodenoscopes, Olympus America, Inc., Fujifilm Medical Systems, USA, Inc., and Hoya Corp. to conduct studies to better understand how their devices were being reprocessed in healthcare settings. The order came after several confirmed reports of infections caused by improperly cleaned or faulty duodenoscopes.
Fujifilm Medical Systems issued revised reprocessing instructions late last month for Model ED-530XT duodenoscopes, according to a safety communication issued by the Food and Drug Administration (FDA). The instructions require exacting pre-cleaning, manual cleaning, and high-level disinfection procedures.
While these revised reprocessing instructions are for Model ED-530XT duodenoscopes, the FDA also is encouraging healthcare facilities that use Fuji’s 250 and 450 duodenoscope models to use the revised reprocessing instructions until the revised instructions for those models are final.
Key changes to the reprocessing procedure for Fuji’s ED-530EXT duodenoscope are:
• Pre-cleaning
o During immersion of the scope tip in detergent solution, move the forceps elevator back and forth and aspirate detergent solution while the forceps is raised and while lowered.
• Manual cleaning
o Additional brushing of the distal tip, forceps elevator and elevator recess first using the existing Fujifilm valve cylinder cleaning brush (Model WB11002FW2) and then using the new disposable (Model WB1318DE) cleaning brush.
o Additional flushing of detergent and rinse water onto the forceps elevator/recess while the elevator is both raised and lowered.
o Additional flushing steps and increased channel flushing volumes of detergent and rinse water
• Manual high-level disinfection
o Additional flushing of disinfectant and rinse water onto the forceps elevator/recess while the elevator is both raised and lowered. Additional raising and lowering of the elevator while immersed in disinfectant solution and rinse water.
o Additional flushing steps and increased flushing volumes of disinfectant and rinse water through the duodenoscope’s internal channels.
For more information visit the FDA website.
Want to receive articles like this one in your inbox? Subscribe to Accreditation Insider!
Related Products
Most Popular
- Articles
-
- Don't forget the three checks in medication administration
- Note similarities and differences between HCPCS, CPT® codes
- The consequences of an incomplete medical record
- OB services: Coding inside and outside of the package
- Complications from immobility by body system
- Q&A: Primary, principal, and secondary diagnoses
- Practice the six rights of medication administration
- Nursing responsibilities for managing pain
- Differentiate between types of wound debridement
- Skills of effective case managers
- E-mailed
-
- Correctly bill ancillary bedside procedures in addition to the room rate
- Q/A: Coding infusions to correct low potassium levels
- Q&A: Utilization Review Committee Membership
- Q&A: Bill blood administration the same way for inpatient and outpatient accounts
- Q&A: A second look at encephalopathy as integral to seizures/CVA
- OB services: Coding inside and outside of the package
- Know the medical gas cylinder storage requirements
- Intravenous therapy guidelines
- Coding, billing, and documentation tips for teaching physicians, interns, residents, and students
- Coding tip: Watch for different codes for SI joint injections
- Searched